Background:
Coronary artery fistula is a rare entity of cardiovascular anomaly. The etiology is often congenital due to malformation of coronary artery vessels. They are mostly asymptomatic, while some cases may be present with complications such as congestive heart failure, myocardial infarction, or pulmonary hypertension. This case study aims to evaluate the Successful coronary artery fistula closure using Amplatzer duct occluder II device in a 10-months-old infant.

Case Presentation: The 10-month-old female infant was present with a coronary artery fistula diagnosed at three months old. The echocardiography revealed a right coronary artery fistula to the right atrium with an ampulla size of 7.29 mm, waist of 3.11 mm, and length of 4.71 mm. The patient underwent cardiac catheterization and coronary artery fistula closure with a transvenous technique approach using Amplatzer duct occluder II. The procedure was successful, and the echocardiography examination found no residual flow from the fistula. This is the first reported case of successful coronary artery fistula closure using the Amplatzer duct occluder II device.

Conclusion: Fistula closure using the Amplatzer Duct Occluder II device, in this case, is preferred because it is safe and efficient to use in small infants due to its small sheath delivery catheter caliber.

References

  1. Torres S, Vasconcelos M, Silva MT, Moreira J, Silva JC, Macedo F. Coronary artery fistulas: A 12-year single-center experience. Revista Portuguesa de Cardiologia. 2022;41(10):843-850.
  2. Vindhyal M, Kolli L, Kannekanti N, Boppana V. Atypical Presentation of Coronary Artery Fistula: Case Report and Review of the Literature. Cureus. 2020;12(4):e7735.
  3. Acitelli A, Bencivenga S, Giannico MB, Lanzillo C, Maresca L, Petroni R, et al. Coronary artery fistula diagnosed by echocardiography during NSTEMI: case report and review of literature. Case Reports in Cardiology. 2019;1(1):1-5.
  4. Reddy G, Davies JE, Holmes DR, Schaff HV, Singh SP, Alli OO. Coronary artery fistulae. Circulation: Cardiovascular Interventions. 2015;8(11):e003062.
  5. Costa RND, Ribeiro MS, Silva AFD, Ribeiro RA, Berwanger O, Biasi A, et al. Incremental Cost-Effectiveness of Surgical vs. Percutaneous Treatment of Patent Ductus Arteriosus with the Amplatzer TM Duct Occluder in Children: A Systematic Review. Revista Brasileira de Cardiologia Invasiva. 2014;22(1):168-179.
  6. Shahanavaz S, Winlaw DS, Opotowsky AR. What is blocking transcatheter ventricular septal defect closure?. Journal of the American Heart Association. 2022;11(7):e024963.
  7. Cen H, Peng B, Li J, Chen S, Sun P. Efficacy and safety of the Amplatzer Duct Occluder II for ventricular septal defect closure: a meta-analysis. Kardiologia Polska (Polish Heart Journal). 2021;79(4):401-409.
  8. Nepal S, Annamaraju P. Coronary Arteriovenous Fistula. StatPearls. 2022. Available at: https://www.ncbi.nlm.nih.gov/books/NBK554618/. Accessed on 20th March 2023.
  9. Buzaev IV, Plechev VV, Khalikova G, Khabirova K, Nikolaeva IE, Onorato EM. Case Report: Coil Occlusion of Two Congenital Coronary Cameral Fistulas Connecting Right and Left Circumflex Arteries to the Right Ventricle: An Innovative Stent-Assisted Technique. Frontiers in Cardiovascular Medicine. 2022;8(1):2086.
  10. Yun G, Nam TH, Chun EJ. Coronary artery fistulas: pathophysiology, imaging findings, and management. Radiographics. 2018;38(3):688-703.
  11. Tzeng PR, Lee KC, Wei J, Chuang YC, Chang CY. Large coronary artery fistula. Formosan Journal of Surgery. 2015;48(4):140-143.
  12. Ghumman GM, Khan A, Shafqat M, Shah J, Singh H. A Coronary Cameral Fistula Associated With Incessant Ventricular Arrhythmias. Cureus. 2023;15(3):1-4.
  13. Al-Hijji M, El Sabbagh A, El Hajj S, AlKhouli M, El Sabawi B, Cabalka A. Coronary artery fistulas: indications, techniques, outcomes, and complications of transcatheter fistula closure. Cardiovascular Interventions. 2021;14(13):1393-1406.
  14. Raju MG, Goyal SK, Punnam SR, Shah DO, Smith GF, Abela GS. Coronary artery fistula: a case series with review of the literature. Journal of Cardiology. 2009;53(3):467-472.
  15. Challoumas D, Pericleous A, Dimitrakaki IA, Danelatos C, Dimitrakakis G. Coronary arteriovenous fistulae: a review. International Journal of Angiology. 2014;23(01):001-010.
  16. Gowda ST, Latson L, Sivakumar K, Hiremath G, Crystal M, Law M, et al. Anatomical classification and posttreatment remodeling characteristics to guide management and follow-up of neonates and infants with coronary artery fistula: a multicenter study from the Coronary Artery Fistula Registry. Circulation: Cardiovascular Interventions. 2021;14(12):e009750.
  17. Rao S, Agasthi P. Coronary Artery Fistula. StatPearls. 2022. Available at: https://pubmed.ncbi.nlm.nih.gov/32644617/. Accessed on: 20th March 2023.
  18. Koneru J, Samuel A, Joshi M, Hamden A, Shamoon FE, Bikkina M. Coronary anomaly and coronary artery fistula as the cause of angina pectoris with literature review. Case reports in vascular medicine. 2011;1(1):1-6.
  19. Gür AK, Karakurt A, Yk K. Congenital coronary-pulmonary artery fistula originating from right and left coronary artery. Dicle Med J/Dicle Tıp Dergisi. 2013;40(1):659-662.
  20. Prabawa IPY, Lestari AAW, Muliarta IM, Mardhika PE, Pertiwi GAR, Bhargah A, et al. The Stromal Cell-derived Factor-1/CXCL12 3’A-gene Polymorphism is Related to the Increased Risk of Coronary Artery Disease: A Systematic Review and Meta-analysis. Open Access Macedonian Journal of Medical Sciences. 2022;8(F):197-202.
  21. Wismiyarso DE, Adriana C, Mangkoesoebroto AP, Christiawan A, Sulma AN, Effendi LHDP. Total occlusion of coronary artery without ST-segment elevation a case series of ‘de Winter’ electrocardiogram pattern. Bali Medical Journal. 2021;10(1):347–350.
  22. Fitrasanti BI, Onggo A, Sugirman W, Ciptawan B. Prevalence of cutaneous markers in coronary artery disease cases. Bali Medical Journal. 2021;10(2):877–880.
  23. Gunawijaya E, Yantie NPVK, Gunawijaya AP. Results and cost analysis of Amplatzer® Ductal Occluder-ii, Nit Occlud® Lê Coil, and Konar-MF® ventricular septal defect occluder for percutaneous ventricular septal defect closure. Bali Medical Journal. 2023;12(1):1054–1058.
  24. Venczelova Z, Tittel P, Masura, J. The new Amplatzer duct occluder II: when is its use advantageous?. Cardiology in the Young. 2011;21(5):495-504.
  25. Pepeta L, Dippenaar A. Ductal closure using the Amplatzer duct occluder type two: Experience in Port Elizabeth hospital complex, South Africa: Cardiovascular topic. Cardiovascular Journal of Africa. 2013;24(6):202-207.
  26. Baspinar O, Irdem A, Sivasli E, Sahin DA, Kilinc M. Comparison of the efficacy of different-sized Amplatzer duct occluders (I, II, and II AS) in children weighing less than 10 kg. Pediatric cardiology. 2013;34(1):88-94.
  27. Arora H, Aggarwal N, Agarwal M, Joshi RK, Joshi R. Device closure of aortopulmonary window using ADO II in an infant. IHJ Cardiovascular Case Reports (CVCR). 2018;2(3):211-213.
  28. Behera SK, Danon S, Levi DS, Moore JW. Transcatheter closure of coronary artery fistulae using the Amplatzer Duct Occluder. Catheterization and cardiovascular interventions. 2006;68(2):242-248.

How to Cite

Faisal, A., Hidayat, T., Rahman, M. A., & Umayasa, A. (2023). Successful coronary artery fistula closure using Amplatzer duct occluder II device in a 10-months-old infant. Bali Medical Journal, 12(2), 2284–2288. https://doi.org/10.15562/bmj.v12i2.4274